For acceptance testing and routine quality control measurements on diagnostic X-ray units.
NOMEX
Key Features
- True Precision: Dose accuracy within the RAD/FLU range typically ± 1.5%, within the MAM range ± 2.5% and kV accuracy within the RAD/FLU range typically ± 0.75% or ± 0.5 kV
- Angular independent for positioning within the beam
- Ergonomic housing avoids shifting after positioning
- Automatic start/stop and auto reset functions
- Automatic dose and kV ranging avoids double exposures
- Automatic total filtration compensation up to 40 mm Al
- Automatic HVL detection for all applications including mammography
Hardware
- Miniaturized multi-parameter measuring device compliant with IEC 61674
- Has passed the Type Approval Under German Law as a diagnostic dosemeter by the German National Metrology Institute, PTB
- For measurements in RAD/FLU, DENT, CT, MAM (incl. Mo/Mo, Mo/Al, Mo/Rh, W/Al, W/Rh, W/Ag, Rh/Rh)
- Connection of the CE marked class IIb certified multimeter to a PC via USB
- Single exposure simultaneously captures: dose, dose rate, dose per pulse, pulses, frequency, irradiation time, tube voltage (kVpmax, kVpmean, PPV) , total filtration, half value layer, together with waveforms for both kV and dose rate
- Practical peak voltage (PPV) calculation according to the algorithm of IEC 61676
- Sensitivity dose measurements starting from 100 nGy/s
Software
- Modifiable parameter displaying (view all parameters at a glance or just selected ones)
- Statistic calculations (mean value, standard deviation, coefficient of variation) by simply marking the results of interest
- Adjustable timer functions (start or stop delays) and waveform sampling up to 60 seconds
- Data and waveform export as XML or CSV file for subsequent data processing in Excel
- Software menu available in different languages (German, English, French, Spanish, Portuguese, Italian, Russian, Chinese, Japanese)