Roos® Electron Chamber

Klicken um Bild zu vergrößernRoos Electron Chamber

Precision plane parallel chamber for absolute dosimetry of high-energy electron radiation in water and solid state phantoms

  • Vented sensitive volume of 0.35 cm3
  • Reference chamber for precise absolute electron dosimetry
  • Perturbation-free design and minimal polarity effect
  • Suitable for use in water and in solid state phantoms

The Roos electron chamber1 is used as a reference electron chamber. It is recommended by the IAEA2 for high precision electron dosimetry in radiation therapy. The chamber has a 4 mm wide guard ring to exclude any perturbation effect even at low electron energies. The polarity effect is negligible (< 0.5 % at 10 MeV). The energy response is only influenced by the stopping power ratios water / air. The chamber is waterproof for absolute dose and depth dose measurements in a water phantom. The coated acrylic entrance window has a thickness of 1.1 mm. The nominal useful energy range is from 2 MeV to 45 MeV. A calibration certificate with a 60Co calibration factor given in absorbed dose to water is included. Air density correction is required for each measurement. A radioactive check device with adapter is available as an option. The chamber cable length is 1.08 m.

1 The Roos electron chamber was developed in cooperation with Dr. Roos, PTB-Braunschweig, German Federal Institute of Physics and Metrology (National Laboratory of Germany)
2 Technical Report No. TRS-381. The Use of Plane Parallel Ionization Chambers in High Energy Electron and Photon Beams, IAEA (International Atomic Energy Agency), Vienna 1997


Product Information

 

Cookies at PTW websites

At our websites we use cookies to collect statistic information in order to optimise the use of our website. By proceeding further into our website, you accept our use of cookies. Read more about our cookie policy and instructions on how to delete cookies.
Accept cookies